Three Fogging Test Methods for Automotive’s Inner Decoration

This paper introduces three fogging testing methods for automotive inner decoration.
By: Shirley Wu
 
July 7, 2009 - PRLog -- Abstract: The paper introduces three fogging test methods of auto motive's inner decoration, stressing the fogging lab's requirements for the environment, basic facility, testing equipment, parts of instrument and the operators. In addition to that, it emphasizes the main points of constructing fogging lab.
 Keywords: fogging test, environment, basic facility, instruments, staffs

Overview
 Through fogging test to materials of auto's inner decoration, volatile substances of those materials can be controlled so as to effectively lower the environmental contamination inside cars. Therefore, fogging test has become an important means for automotive manufacturers as well as auto's inner decoration manufacturers to control product quality.
 At present, relevant domestic and international fogging test standards for auto's inner decoration altogether involve three kinds of test methods: the reflectometric method, the gravimetric method and the haze method. These standards and their relevant testing methods are as follows:
      DIN 75201-1992-Reflectometric method, gravimetric method
      SAE J1756-1994- Reflectometric method, gravimetric method
      ISO 6452-2000- Reflectometric method, gravimetric method
      NES M0161- Haze method
      ES-X83231- Haze method, gravimetric method
      QB/T 2728-2005- Reflectometric method, gravimetric method

2. Fogging Device
 The figure below is the structural representation of fogging device in fogging test. A sample is placed under specimen presser at the bottom of a glass beaker, which is immersed in a constant temperature oil bath. The beaker is covered with oil and higher temperature resistant fluorine seal ring, with fogging carrier (position G), filter paper, cooling plate placed above sequentially. The substances emitted from the leather sample under heat condense on the cooling plate.
 Except for position G, test equipments for these three methods are the same. For reflectometric and hazing method, the G part is a fogging glass plate. Test results are obtained through measuring the opaqueness or fogging value of the glass plate before and after test. In Gravimetric method, position G is an aluminum foil with a little bigger circular glass plate 3 mm in thickness placed above. Test results of this method are obtained through measuring the quantity of condensate that formed on a cold surface.

3. Environmental Requirements
 So long as constant high temperature bath and cooling plate can work under required temperature, fogging process of the test does not require high on external environment. Environment has no effects on it fundamentally. However, fogging carriers, such as aluminum foil or glass plate, have strict requirements on environment and even the slightest change will produce great influence.
 In Almost all standards, fog-formed glass plate is required to be placed under a temperature of 23℃±2℃ and conditioned under the humidity of 50%±5%RH in a windless and sunless environment. Fogging carrier has some condensed water inevitably. Environmental temperature and humidity, which are difficult to be controlled efficiently, are of vital importance to water evaporation. The uncertainty of environmental control will directly affect accuracy of final testing results.
In standards DIN 75201, ISO 6452 and QB/T 2728, it is specified that fog-formed aluminum foil should be conditioned in desiccators. In this case, users only need to take environmental temperature into condensation and need not consider humidity and gas flow any more. Consequently, conditioning environment of fog-formed aluminum foil is more easily controlled.

 4. Test Results Calibration
 Since fogging test is complex in operation procedures and is easily subject to environment variation, the three standards, DIN 75201, ISO 6452 and QB/T 2728, provide the same calibration method to evaluate the correctness of final test results.
 For reflectometric method, 10±0.1g DIDP are put into the beaker that is immersed in 100℃ constant temperature bath; the cool plate is 21℃; and the final result should be within 77±3%. Otherwise, the test is regarded as unsuccessful. Users should find out the reason and re-test.
 In gravimetric method, 10±0.1g DOP are put into the beaker that is immersed in 100℃ constant temperature bath; the cool plate is 21℃, and the final results should be within 4.9±0.25mg. Otherwise the test is regarded as unsuccessful. Users should Find out the reason and retest.
 For haze method, no calibration method has been provided in the above-mentioned standards.

5. Comparison of the Three Methods
 The table below is a comparison of the items of the three methods
 5.1 Devices to be cleaned:
  5.1.1 Reflectometric method: Beaker, specimen presser, seal ring, glass plate
  5.1.2 Gravimetric method: Beaker, specimen presser, seal ring
  5.1.3 Haze method: Beaker, specimen presser, seal ring, glass plate
 5.2 Fogging carrier:
  5.2.1 Reflectometric method: Glass plate
  5.2.2 Gravimetric method: Aluminum foil
  5.2.3 Haze method: Glass plate
 5.3 Environmental requirements
  5.3.1 Reflectometric method: Temperature 23℃±2℃,humidity 50%±5%RH
  5.3.2 Gravimetric method: Temperature 23℃±2℃
 5.3.3 Haze method: Temperature 23℃±2℃, humidity 50%±5%RH
 5.4 Calibration agent
  5.4.1 Reflectometric method: DIDP
  5.4.2 Gravimetric method: DOP
  5.4.3 Haze method: No agent
5.5 Calibration  period
  5.5.1 Reflectometric method: 4 hours
  5.5.2 Gravimetric method: 20 hours
  5.5.3 Haze method: 5hours
 5.6 Success rate of test
  5.6.1 Reflectometric method: About 75%
  5.6.2 Gravimetric method: About 97%
  5.6.3 Haze method: not mentioned

 With its statistics and analysis to 19 reflectometric method tests, 32 gravimetric tests and 9 hazing tests, Labthink Fogging Lab has obtained the value of success rate of this table. Success rate for reflectometric method is usually within 69%~74%, a little lower to the calibration range. For the same instruments using the same operating method in the same external environment, almost all the test using gravimetric method are successful, which completely indicates the better operability of gravimetric method.

6. Conclusion
 The reflectometric method and hazing method are similar to each other and both are used to test optical property of glass plate. These two methods are adopted when users attach great importance to the impairing visibility caused by volatile substance from auto's inner decoration condensed on windscreen. While in adopting gravimetric method, users pay more attention to the influence of hazardous volatile substances on human health.

Further information please feel free to contact Shirley Wu via trade2@labthink.cn, skype:shirley-labthink, MSN:shirleywu_labthink@hotmail.com or (tel)0086-531-85068709.

# # #

Labthink provide most excellent quality control solutions for pharmaceutical, foodstuff, cosmetics, packaging, printing, adhesive, automotive, petrochemistry, environment, biology, new energy, construction, aviation and electronic industries worldwide.
End
Source:Shirley Wu
Email:***@labthink.cn Email Verified
Zip:250031
Industry:Technology
Location:jinan - Shandong - China
Account Email Address Verified     Disclaimer     Report Abuse
Labthink Instruments Co., Ltd. PRs
Trending News
Most Viewed
Top Daily News



Like PRLog?
9K2K1K
Click to Share