Adhesive Bandage Peeling Testing Methods and Applications

The peeling test of adhesive bandage is of crucial importance.
By: Labthink Instruments Co.,Ltd
 
April 9, 2009 - PRLog -- Peeling strength of adhesive bandage is of crucial importance.
According to pharmacopoeia, 180 degree peeling strength testing is adopted. Take Labthink BLD-200S Peeling Strength Tester or XLW Auto Tensile Tester as example.
Stick the back side of the adhesive bandages on the testing panel with double-side adhesive tape. Strengthen with adhesive tape so that sample can be even on testing panel. Place clean polymer film on adhesive side of the sample, and roll three rounds with 2000g roller on the sample to avoid bubbles. Lay the sample aside for 20 to 40 minutes before test.  
Bend the polymer film and clamp the free end of the film and testing panel on the tester. Peeling side must be in linear direction with the tester. Set testing speed at 300mm/min±10mm/min and start. The results of peeling strength are with the units of KN/m or N/mm.
Labthink is willing to have further exchanges with adhesive bandage manufacturers in the peeling strength testing and quality control.

Labthink Instruments Co., Ltd.
No. 144 Wuyingshan Road, Jinan 250031, China
Tracy Bao  info@labthink.cn
Tel: 0086 531 85061153  fax:0086 531 85812140

# # #

We Labthink provide a wide variety of testing instruments, services and consultation for industries of flexible packaging, printing, adhesive and automotive nonmetal.
End
Source:Labthink Instruments Co.,Ltd
Email:***@labthink.cn Email Verified
Zip:250031
Tags:Adhesive Bandage, Adhesive Bandage Peeling, Adhesive Bandage Peeling Testing
Location:jinan - Shandong - China
Account Email Address Verified     Disclaimer     Report Abuse
Labthink Instruments Co., Itd. PRs
Trending News
Most Viewed
Top Daily News



Like PRLog?
9K2K1K
Click to Share