Validation of the Memmert hot-air steriliser

Process validation for a Memmert hot-air sterilisers is no problem at all. As an example, the Schwabach manufacturer of temperature control appliances commissioned the independent and accredited test laboratory biomedis to validate two appliances.
By: falkner@mond-online.com
 
Nov. 29, 2010 - PRLog -- Updated November 2010

Process validation for a Memmert hot-air sterilisers is no problem at all. As an example, the Schwabach manufacturer of temperature control appliances commissioned the independent and accredited test laboratory biomedis to validate two appliances. The test setup was based on the international standard ISO 20857 for the validation of sterilisation processes by means of hot air, published in August 2010.

In order to protect patients, the requirements made on the quality of medical products are extremely strict, and quite rightly so. For this reason, it was important for Memmert to have the absolute reliability of its hot-air sterilisers verified by an independent organisation. The test setup was based on the guidelines of the DGKH (German Association for Hospital Hygiene) and on the specifications of the standard ISO 20857 for the sterilisation of medical products. The test was performed once without a load, once with mixed load and twice at full load, with some metallic instruments partially wrapped in plastic or aluminium foil as well as glass containers. In the SNE 200 steriliser (with a chamber volume of 32 litres), a total of 16 measurement sensors monitored the constant temperature. Due to the larger volume of 108 litres in the SFP 500 steriliser, 27 sensors were distributed in the chamber and attached directly on the chamber load. The results were identical for both appliances. During each measurement, the minimum temperature of 180 °C was reached and maintained for at least 30 minutes at each measuring point.

Setpoint-dependent sterilisation time
Apart from optimum temperature distribution and temperature stability in the working chamber, a third, crucial requirement is made of hot-air sterilisers: It must be possible to adjust the sterilisation time depending on when the setpoint temperature is reached. In the Memmert hot-air steriliser, the SPWT function, the continuation of the programme depending on the setpoint being reached, ensures that sterilisation does not begin before the predefined sterilisation temperature has been reached. In addition to the standard Pt100 temperature sensor, up to three external temperature sensors can be integrated in the Memmert hot-air steriliser programme sequence and thus ensure that the sterilisation time does not start before the set temperature has been reached at all the measurement points.


Responsible for content
Angelika Henneberg
Memmert GmbH + Co. KG
Äußere Rittersbacher Str. 38
91126 Schwabach, Germany
Telefon +49 (0)9122 925-0
ahenneberg@memmert.com
www.memmert.com

# # #

Memmert develops and manufactures laboratory appliances. The product range includes devices such as heating oven, incubator, sterilizer, vacuum drying oven, Peltier-cooled incubator, CO2 incubator, humidity chamber, constant climate chamber and much more.
End
Source:falkner@mond-online.com
Email:***@memmert.com
Zip:91126
Tags:Laboratory Appliance, Drying Oven, Incubator, Cooled Incubator, Storage Chamber, Hot-air Steriliser, Humidity Chamber
Industry:Industrial, Technology, Environment
Location:Schwabach - Bavaria - Germany
Account Email Address Verified     Disclaimer     Report Abuse
Memmert GmbH + Co. KG News
Trending
Most Viewed
Daily News



Like PRLog?
9K2K1K
Click to Share